GAV - recommended pressure levels

The GAV is available in six different pressure configurations. It is MRI-safe and calibrated in accordance with strict quality control standards.

recommended pressure levels GAV (pdf)

 
   
 
recommended pressure levels        
The required opening pressure of the GAV depends on the individual patient's age, mobility, height and the body shape:        
Age
Mobility
Height
Body shape
recommended pressure level
in cmH2O
older than 60 years
mobile
less than 1,60 m
adipose
4 / 24 paediGAV
slim
5 / 30
greater than 1,60 m
adipose
5 / 30
slim
5 / 35
immobile
less than 1,60 m
adipose
4 / 24 paediGAV
slim
4 / 24 paediGAV
greater than 1,60 m
adipose
4 / 24 paediGAV
slim
5 / 30
younger than 60 years
mobile
less than 1,60 m
adipose
4 / 24 paediGAV
slim
5 / 30
1,60 m to 1,80 m
adipose
5 / 30
slim
5 / 35
greater than 1,80 m
adipose
5 / 30
slim
5 / 40
immobile
smaller than 1,60 m
adipose
4 / 24 paediGAV
slim
5 / 30
greater than 1,60 m
adipose
4 / 24 paediGAV
slim
5 / 30

 

         

Reading the pressure setting from an X-ray image

Every single pressure configuration is individually coded and can be identified on the GAV in an X-ray image post-operatively.
Fig.1 shows the X-ray image of a GAV
(10 / 40 cmH2O).

 
Fig. 1  
         
The GAV is available in the following pressure settings:        
opening pressure horizontal / vertical position (cmH2O)
coding GAV in X-ray image
 
 
5 / 30
5 / 35
5 / 40
10 / 30
10 / 40