DSV ® Pressure setting

The DUALSWITCH-VALVE is available in different pressure configurations.

Different pressure settings are available either for the low pressure chamber (horizontal body position of the patient) or the high pressure chamber (vertical body position). The recommended pressure level for the low pressure chamber is dependent on the origin of the hydrocephalus.

The height of the patient, however, is the determining factor when considering the opening pressure required for the high level pressure chamber.

 

 

recommended pressure levels

HightStandard opening pressure horizontal / vertical (cmH2O)NPH valve opening pressure horizontal / vertical (cmH2O)
Patients shorter than 1,60 m 10 / 305 / 30
Patients from 1,60 m to 1,80 m 10 / 405 / 40
Patients taller than 1,80 m 10 / 505 / 50

Reading the pressure setting from an X-ray image

The picture on the right shows the X-ray image of a DSV (pressure setting 13/30 cmH2O).

 

In addition to the standard pressure settings the DUALSWITCH-VALVE is also available in the following special configurations:

opening pressure horizontal / vertical (cmH2O)coding DSV in X-ray imageopening pressure
horizontal / vertical (cmH2O)
coding DSV in X-ray image
5 / 30Kodierung DSV im Röntgenbild mit Öffnungsdruck 5/3013 / 30Kodierung DSV im Röntgenbild mit Öffnungsdruck 13/30
5 / 40Kodierung DSV im Röntgenbild mit Öffnungsdruck 5/4013 / 40Kodierung DSV im Röntgenbild mit Öffnungsdruck 13/40
5 / 50Kodierung DSV im Röntgenbild mit Öffnungsdruck 5/5013 / 50Kodierung DSV im Röntgenbild mit Öffnungsdruck 13/50
10 / 30Kodierung DSV im Röntgenbild mit Öffnungsdruck 10/3016 / 30Kodierung DSV im Röntgenbild mit Öffnungsdruck 16/30
10 / 40Kodierung DSV im Röntgenbild mit Öffnungsdruck 10/4016 / 40Kodierung DSV im Röntgenbild mit Öffnungsdruck 16/40
10 / 50Kodierung DSV im Röntgenbild mit Öffnungsdruck 10/5016 / 50Kodierung DSV im Röntgenbild mit Öffnungsdruck 16/50